Test examination
RPO THK-basic
Typical values of the equivalent dos rate (in mSv per mA s) at a distance of 1 meter from the focus of the X-ray tube, for different values of tube voltage and filter thickness.
tube voltage | thickness aluminum filter | ||
---|---|---|---|
1 mm | 2 mm | 3 mm | |
50 kV | 0.07 | 0.05 | 0.03 |
70 kV | 0.15 | 0.09 | 0.05 |
90 kV | 0.20 | 0.13 | 0.08 |
Last modified: | 12 November 2020 11.54 a.m. |
View this page in: Nederlands